Z-1837-2019 Class II Terminated
Recall Details
- Product Type
- Devices
- Report Date
- June 26, 2019
- Initiation Date
- April 15, 2019
- Termination Date
- October 5, 2020
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 28 units
Product Description
CIVCO Solstice(TM) SRS Immobilization System, Part Numbers: a) CHS03, b) CHS04 - Product Usage: The Tilting Head Fixation (Solstice) device is comprised of a baseplate and bowl with a locking lever. The bowl is permanently attached to the baseplate and the locking level allows the bowl to be tilted up to approximately 10¿ in relation to the baseplate at infinite intervals. The baseplate is attached to the treatment or simulation couch top or overlay by pins. When used with customizable cushion and thermoplastic mask, the system can immobilize the patient to under 1 millimeter. The provider can correct the patient position for simulation and treatments by tilting the device.
Reason for Recall
There is a potential for movement of the Solstice Tilting Head Fixation System during patient setup and/or treatment.
Distribution Pattern
Worldwide distribution - US Nationwide distributed in the states of CA, HI, PA, KY, FL, WI, AK and countries of Australia, Canada, Italy, Netherlands, Switzerland, Peru.
Code Information
Part Numbers: a) CHS03, UDI 00841439112489 Lot Numbers: M602380, M680480, M847590, M870360 b) CHS04, UDI 00841439112496 Lot Numbers: M680530, M789100, M789110, M839690, M938500