Z-2253-2014 Class II Terminated
Recalled by Ortho Development Corporation
— Draper, UT
FDA device recall Z-2253-2014 was initiated by Ortho Development Corporation on Jul 11, 2014 and is designated Class II. Reason for recall: Ortho Development is recalling various Pivot Bipolar Cup, Patella, PS Tibial Insert, UC Tibila Insert, and CK Tibil Insert because the foil pouches did not properly seal, allowing air to seep back… The recall status is terminated (terminated Oct 6, 2014). Affected quantity: 188.
Recall details
- Product Type
- Devices
- Report Date
- Aug 27, 2014
- Initiation Date
- Jul 11, 2014
- Termination Date
- Oct 6, 2014
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 188
Product description
Pivot Bipolar Cup 22x42/43-51/53mm, REF 133-YYYY Product Usage: Used in a hemiarthroplasty procedure as the bearing surface to articulate with both the CoCr femoral head and the natural acetabulum
Reason for recall
Ortho Development is recalling various Pivot Bipolar Cup, Patella, PS Tibial Insert, UC Tibila Insert, and CK Tibil Insert because the foil pouches did not properly seal, allowing air to seep back into the pouch after sealing which increases the risk that the material properties of the implant would be substantially degraded due to oxidation.
Distribution pattern
Worldwide Distribution - US Nationwide and the country of Japan. No military/govt/VA consignees.
Code information
Pivot Bipolar Cup 22x42mm,Part 133-2242,Lot A136254,Pivot Bipolar Cup 22x43mm,Part 133-2243,Lot A136256,Pivot Bipolar Cup 22x43mm,Part 133-2243,Lot A136255,Pivot Bipolar Cup 22x44mm,Part 133-2244,Lot A136257,Pivot Bipolar Cup 22x45mm,Part 133-2245,Lot A136258,Pivot Bipolar Cup 22x46mm,Part 133-2246,Lot A136259,Pivot Bipolar Cup 22x46mm,Part 133-2246,Lot A136260,Pivot Bipolar Cup 22x47mm,Part 133-2247,Lot A136261,Pivot Bipolar Cup 22x48mm,Part 133-2248,Lot A136262,Pivot Bipolar Cup 22x49mm,Part 133-2249,Lot A136263,Pivot Bipolar Cup 22x50mm,Part 133-2250,Lot A136265,Pivot Bipolar Cup 22x51mm,Part 133-2251,Lot A136266,Pivot Bipolar Cup 22x53mm,Part 133-2253,Lot A136269,Pivot Bipolar Cup 22x53mm,Part 133-2253,Lot A136268,Pivot Bipolar Cup 22x53mm,Part 133-2253,Lot A136267,Pivot Bipolar Cup 22x53mm,Part 133-2253,Lot A136270,
Source
- Dataset
- openFDA device recall enforcement reports
- FDA record
- This recall in openFDA (JSON)
- Enforcement Report
- Search FDA's Enforcement Report for Z-2253-2014
- openFDA file date
- Nov 19, 2026
- Loaded by RegDataLab
- Nov 20, 2026
These dates say when openFDA published the data file and when RegDataLab last loaded it.
Need the full FDA record for Ortho Development Corporation? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.
Get a free API key
How the connector works