Z-1903-2020 Class II Ongoing

Recalled by Ortho Development Corporation — Draper, UT

Recall Details

Product Type
Devices
Report Date
May 13, 2020
Initiation Date
May 30, 2019
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
27

Product Description

Tibial Tray Nonporous A Size 2; Model # PN 162-1200-A; Material: Ti6Al4V.

Reason for Recall

The locking mechanism of the tibial tray exhibited a manufacturing defect in which a uniform ridge was cut along the inner wall of the "T" and "Y" junctions, which may cause resistance when the Tibial Insert was being assembled.

Distribution Pattern

US: CA, FL, OH, PA, TX, and UT. OUS: Japan.

Code Information

Lot Number: A200590