Z-2253-2016 Class II Terminated

Recalled by The Anspach Effort, Inc. — Palm Beach Gardens, FL

FDA device recall Z-2253-2016 was initiated by The Anspach Effort, Inc. on Jun 15, 2016 and is designated Class II. Reason for recall: The graphics at the attachment interface indicating locked and unlocked position are reversed. The recall status is terminated (terminated Dec 14, 2016). Affected quantity: 18.

Recall Details

Product Type
Devices
Report Date
Jul 27, 2016
Initiation Date
Jun 15, 2016
Termination Date
Dec 14, 2016
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
18

Product Description

Anspach G1 High Speed Electric Handpiece Air Cooled (eG1A) Use for cutting and shaping bone including bones of the cranium and spine.

Reason for Recall

The graphics at the attachment interface indicating locked and unlocked position are reversed.

Distribution Pattern

Worldwide Distribution: US Distribution to states of: GA and FL; and country of: Switzerland.

Code Information

Lot #'s: H29309439804, H44310033104, K09311396703, and K09311396704.