Z-2262-2019 Class II Terminated

Recalled by Degania Silicone, Ltd. — Dganya Bet, N/A

FDA device recall Z-2262-2019 was initiated by Degania Silicone, Ltd. on Jun 28, 2019 and is designated Class II. Reason for recall: 15FR and 19FR round fluted drain channel drain may break during use (mostly during removal) in connection point between clear tube and white profile part. requiring additional surgical intervention… The recall status is terminated (terminated Oct 22, 2020). Affected quantity: 6080.

Recall Details

Product Type
Devices
Report Date
Aug 21, 2019
Initiation Date
Jun 28, 2019
Termination Date
Oct 22, 2020
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
6080

Product Description

Cardinal Health Jackson-Pratt Channel Drain 15FR, Full flutes, with trocar Product Number: JP-2189 An introduction/drainage catheter is a device that is a flexible single or multilumen tube intended to be used to introduce nondrug fluids into body cavities other than blood vessels, drain fluids from body cavities, or evaluate certain physiologic conditions.

Reason for Recall

15FR and 19FR round fluted drain channel drain may break during use (mostly during removal) in connection point between clear tube and white profile part. requiring additional surgical intervention to remove the broken end from patient's body

Distribution Pattern

Nationwide Distribution

Code Information

Lot Codes: P1841752 P1852268 P1845823

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-2262-2019
openFDA file date
Oct 23, 2026
Loaded by RegDataLab
Oct 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.