Z-2264-2025 Class II Ongoing

Recalled by Beckman Coulter Inc. — Brea, CA

FDA device recall Z-2264-2025 was initiated by Beckman Coulter Inc. on Jun 26, 2025 and is designated Class II. Reason for recall: Beckman Coulter identified that the Unsaturated Iron Binding Capacity (UIBC) assay is not meeting their labeled hemoglobin interference claims of no significant interference (i.e., less than 10%) up… The recall status is ongoing. Affected quantity: 56,696 units.

Recall details

Product Type
Devices
Report Date
Aug 13, 2025
Initiation Date
Jun 26, 2025
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
56,696 units

Product description

UIBC (Unsaturated Iron Binding Capacity), REF: OSR61205,

Reason for recall

Beckman Coulter identified that the Unsaturated Iron Binding Capacity (UIBC) assay is not meeting their labeled hemoglobin interference claims of no significant interference (i.e., less than 10%) up to 200 mg/dL hemoglobin as indicated in their instructions for use. Initial internal testing confirmed that samples with low UIBC levels failed the claimed hemolysis interference specifications with reported biases up to -43.6 % UIBC when hemolyzed serum samples contained 200 mg/dL hemoglobin.

Distribution pattern

US: AK, AL, AR, CO, CT, DC, DE, ID, UT OUS: Algeria, Argentina, Australia, Bahrain, Bangladesh, Belarus, Belgium, Bosnia and Herzegovina, Brazil, Bulgaria, Canada, China, Colombia, Costa Rica, Croatia, Czechia, Egypt, El Salvador, Ethiopia, Georgia, Germany, Ghana, Gibraltar, Greece, Hong Kong, Hungary, India, Indonesia, Ireland, Israel, Italy, Jordan, Kazakhstan, Korea, Republic of, Kuwait, Lebanon, Libya, Lithuania, Malaysia, Maldives, Mexico, Morocco, Mozambique, Myanmar, Nepal, Netherlands, Nigeria, North Macedonia, Oman, Pakistan, Panama, Peru, Philippines, Poland, Portugal, Puerto Rico, Qatar, Romania, Russian Federation, Saudi Arabia, Serbia, Singapore, Slovakia, Somalia, South Africa, Spain, Sri Lanka, Sweden, Taiwan, Province of China, Thailand, Tunisia, Turkey, United Arab Emirates, United Kingdom of Great Britain and Northern Ireland, Uzbekistan, Viet Nam, Yemen.

Code information

All Lots/(01)15099590011925

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-2264-2025
openFDA file date
Nov 16, 2026
Loaded by RegDataLab
Nov 17, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Source: openFDA record for Z-2264-2025. Report a data issue: hello@regdatalab.com.