Z-2277-2018 Class II Terminated

Recalled by Philips Medical Systems (Cleveland) Inc — Cleveland, OH

FDA device recall Z-2277-2018 was initiated by Philips Medical Systems (Cleveland) Inc on Mar 28, 2018 and is designated Class II. Reason for recall: A problem has been detected in the X-Ray Tube. The affected tubes have a production defect that allows oil leakage from the tube housing, if it were to re-occur, could pose a risk for patients or… The recall status is terminated (terminated Apr 21, 2020).

Recall Details

Product Type
Devices
Report Date
Jul 4, 2018
Initiation Date
Mar 28, 2018
Termination Date
Apr 21, 2020
Voluntary/Mandated
Voluntary: Firm initiated

Product Description

Ingenuity Core128/Elite China 728324

Reason for Recall

A problem has been detected in the X-Ray Tube. The affected tubes have a production defect that allows oil leakage from the tube housing, if it were to re-occur, could pose a risk for patients or users.

Distribution Pattern

The systems were distributed to the following US states: HI, KS, and MO. The systems were distributed to the following foreign countries: Algeria, Austria, Brazil, China, Indonesia, Japan, Kenya, Korea, Mexico, Poland, Turkey, and Uganda.

Code Information

UDI: (01)00884838059863(21)336856, Serial Number: 336856; UDI: (01)00884838059863(21)336851, Serial Number: 336851; UDI: (01)00884838059863(21)336849, Serial Number: 336849; UDI: (01)00884838059863(21)336852, Serial Number: 336852; UDI: (01)00884838059863(21)336853, Serial Number: 336853; UDI: (01)00884838059863(21)336855, Serial Number: 336855; UDI: (01)00884838059863(21)336854, Serial Number: 336854; UDI: (01)00884838059863(21)336850, Serial Number: 336850

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-2277-2018
openFDA file date
Oct 26, 2026
Loaded by RegDataLab
Oct 27, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.