Z-0213-2020 Class I Terminated
FDA device recall Z-0213-2020 was initiated by Philips Medical Systems (Cleveland) Inc on Sep 19, 2019 and is designated Class I. Reason for recall: An issue with the Detector for the Forte Family of cameras may result in either detector 1 or detector 2 falling unimpeded vertically to the end stops of its travel limit. The recall status is terminated (terminated Apr 5, 2021). Affected quantity: 1167 in total.
Recall details
- Product Type
- Devices
- Report Date
- Nov 13, 2019
- Initiation Date
- Sep 19, 2019
- Termination Date
- Apr 5, 2021
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 1167 in total
Product description
Forte Gamma Camera System is intended to produce images depicting the anatomical distribution of single-photon and positron-emitting radioisotopes within the body for interpretation by medical personnel.
Reason for recall
An issue with the Detector for the Forte Family of cameras may result in either detector 1 or detector 2 falling unimpeded vertically to the end stops of its travel limit.
Distribution pattern
Nationwide Distribution.
Code information
System Code: 882020 All systems