Z-2277-2024 Class II Ongoing

Recalled by American Contract Systems Inc — Tiffin, OH

FDA device recall Z-2277-2024 was initiated by American Contract Systems Inc on May 15, 2024 and is designated Class II. Reason for recall: Ethylene Oxide/Ethylene Chlorohydrin (EO/ECH) residuals for the cast padding component exceeded the residual limits for permanent exposure devices as indicated by ANSI/AAMI/ISO 10993-7/(R)2012. The recall status is ongoing. Affected quantity: 48 units.

Recall Details

Product Type
Devices
Report Date
Jul 17, 2024
Initiation Date
May 15, 2024
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
48 units

Product Description

LOWER EXTREMITY - 247542- Procedure tray Catalog Number: BFLE56X

Reason for Recall

Ethylene Oxide/Ethylene Chlorohydrin (EO/ECH) residuals for the cast padding component exceeded the residual limits for permanent exposure devices as indicated by ANSI/AAMI/ISO 10993-7/(R)2012.

Distribution Pattern

US Nationwide distribution.

Code Information

UDI-DI: 00191072214553 LOT#"s: 943241

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-2277-2024
openFDA file date
Oct 26, 2026
Loaded by RegDataLab
Oct 27, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.