Z-2288-2014 Class II Terminated

Recalled by Zimmer, Inc. — Warsaw, IN

FDA device recall Z-2288-2014 was initiated by Zimmer, Inc. on Jul 1, 2014 and is designated Class II. Reason for recall: The firm discovered that the current Personal Tibial/General Fixed Bracket Trays may not have the specific geometry to accommodate the tibial drill and stop guide with proper position/orientation. The… The recall status is terminated (terminated Jun 9, 2015). Affected quantity: 190 units.

Recall Details

Product Type
Devices
Report Date
Aug 27, 2014
Initiation Date
Jul 1, 2014
Termination Date
Jun 9, 2015
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
190 units

Product Description

PERSONA (TM) The Personalized Knee System Tibial General Instrument Tray, non-sterile The intended use of the sterilization case is to organize, store and protect its contents from damage consistent with typical hospital or surgical center circulation (i.e. between central services and the operating room) and during transport between a Zimmer distributorship and a health care facility by Zimmer distributorship personnel.

Reason for Recall

The firm discovered that the current Personal Tibial/General Fixed Bracket Trays may not have the specific geometry to accommodate the tibial drill and stop guide with proper position/orientation. The drill bracket and drill guide are not the correct geometry to allow for a proper fit in the Persona Fixed Case Kit which could lead to improper sterilization of the tibial drill and stop guide.

Distribution Pattern

Worldwide Distribution - USA including AZ, CA, CO, FL, IL, IN, KS, MI, MN, NC, NJ, NY, OH, OR, PA, TN, TX & VA and Internationally to AUSTRALIA, FRANCE, INDIA, ITALY, and GERMANY.

Code Information

Item Number 00-5907-015-00, Lot Numbers: 56518863 & 56518862

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-2288-2014
openFDA file date
Oct 23, 2026
Loaded by RegDataLab
Oct 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.