Z-2300-2016 Class II Terminated

Recalled by Ge Healthcare It — Barrington, IL

Recall Details

Product Type
Devices
Report Date
August 3, 2016
Initiation Date
February 18, 2013
Termination Date
December 13, 2017
Voluntary/Mandated
N/A
Product Quantity
13

Product Description

Imagecast PACS with Centricity RIS-IC versions prior to 10.6 Update Package 18 Product Usage: is an image management system intended to be used by trained professionals, including but not limited to physicians, nurses and medical technicians. The system is used with general purpose computing hardware to acquire, transmit, process and store images and data throughout a clinical environment. Data and images are acquired through DICOM compliant imaging devices and modalities. Lossy compressed mammographic images and digitized film screen images must not be reviewed for primary image interpretation. Mammographic images may only be interpreted using an FDA approved monitor that offers at least 5 Mpixel resolution and meets other technical specifications reviewed and accepted by the FDA.

Reason for Recall

A software defect was discovered that causes images to be out of context with clinical information.

Distribution Pattern

US Nationwide Distribution.

Code Information

Imagecast PACS 3.5.69, 3.5.86, 3.5.87 and 3.6 with Centricity. RIS-IC 10.6 versions prior to update Package 18.