Z-2739-2016 Class II Terminated

Recalled by Ge Healthcare It — Barrington, IL

Recall Details

Product Type
Devices
Report Date
September 14, 2016
Initiation Date
November 16, 2012
Termination Date
January 11, 2017
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
1,379 units

Product Description

Centricity Enterprise Archive is a software product for receiving, archiving and sending of medical data. Qualified system administrators install, monitor and maintain the system. DICOM devices (e.g. modalities, workstations) communicate with the archive using the DICOM protocol (published by ACR-NEMA). The clinical use is limited to the interaction other systems have with the Centricity Enterprise Archive. Centricity Enterprise Archive is not intended for use in patient vicinity. The intended users for Centricity Enterprise Users service personnel and System administrators, and non clinical users. As of V4.0 XDS enabled systems can communicate with the archive using the XDS and XDS-I profiles (published by IHE).

Reason for Recall

There is a potential for data loss associated with empty path names resulting from HL-7 updates when the cache disk is almost full. In an unlikely event that the cache disk is almost full (<100MB) and HL-7 update events are received, the update action fails and an empty path name is written to the EA database. As a result the instance is not deleted from the storage device but no longer referenced. When trying to retrieve the instance an error is returned.

Distribution Pattern

Nationwide Distribution

Code Information

Version: 3.0.x