Z-2301-2024 Class II Ongoing

Recalled by American Contract Systems Inc — Tiffin, OH

FDA device recall Z-2301-2024 was initiated by American Contract Systems Inc on May 15, 2024 and is designated Class II. Reason for recall: Ethylene Oxide/Ethylene Chlorohydrin (EO/ECH) residuals for the cast padding component exceeded the residual limits for permanent exposure devices as indicated by ANSI/AAMI/ISO 10993-7/(R)2012. The recall status is ongoing. Affected quantity: 28 units.

Recall Details

Product Type
Devices
Report Date
Jul 17, 2024
Initiation Date
May 15, 2024
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
28 units

Product Description

KNEE ARTHROSCOPY-Procedure Kit Catalog Number: SLAR14G

Reason for Recall

Ethylene Oxide/Ethylene Chlorohydrin (EO/ECH) residuals for the cast padding component exceeded the residual limits for permanent exposure devices as indicated by ANSI/AAMI/ISO 10993-7/(R)2012.

Distribution Pattern

US Nationwide distribution.

Code Information

UDI-DI: 00191072198563 LOT#"s: 975241 920241