Z-2302-2020 Class II Terminated

Recalled by Siemens Medical Solutions USA, Inc — Malvern, PA

Recall Details

Product Type
Devices
Report Date
June 17, 2020
Initiation Date
April 30, 2020
Termination Date
September 11, 2020
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
15 US

Product Description

Multi Display Manager in the Artis Systems (Image-Intensified Fluoroscopic X-Ray System) as follows: Artis Q floor Model: 10848280; Artis Q ceiling Model: 10848281; Artis Q biplane Model:10848282; Artis zee ceiling Model: 10094137; Artis zee floor Model: 10094135 Product Usage: Artis is a family of dedicated angiography systems developed for single and biplane diagnostic imaging and interventional procedures including, but not limited to, pediatric and obese patients.

Reason for Recall

Artis systems containing a Multi Display Manager of a specific lot, the power distributor may lead to an increased electrical contact resistance or the interruption of electrical contact, and result in planned procedures to be terminated and performed on an alternative x-ray system

Distribution Pattern

US Nationwide distribution.

Code Information

Serial Numbers: Serial 125253 103340 121526 121521 103335 109621 138311 148418 147237 109620 136555 121520 121525 121518 125254