Z-2309-2018 Class II Terminated
Recall Details
- Product Type
- Devices
- Report Date
- July 4, 2018
- Initiation Date
- May 1, 2018
- Termination Date
- May 16, 2019
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 2
Product Description
Artis zee biplane MN, Material no. 10094143, for angiography and whole body radiographic/fluoroscopic procedures.
Reason for Recall
After the Large Display returns from power save mode, it may not show an image, and stay black without showing an error message although X-ray is still possible. The problem does not occur during an ongoing procedure. If the problem occurs, the system cannot be operated normally. It may be necessary to cancel or restart the treatment or transfer the patient to an alternate or another system.
Distribution Pattern
Worldwide distribution. US nationwide including Puerto Rico, Australia, Austria, Bahrain, Belarus, Belgium, Brazil, Canada, Croatia, Czech Republic, Denmark, Estonia, Finland, France, Germany, Hong Kong, India, Indonesia, Ireland, Israel, Italy, Japan, Luxembourg, Mexico, Myanmar, Netherlands, New Zealand, Norway, China, Paraguay, Philippines, Poland, Portugal, Russia, Saudi Arabia, Singapore, South Africa, South Korea, Spain, Sri Lanka, Sweden, Switzerland, Taiwan, Thailand, Turkey, United Arab Emirates, and United Kingdom.
Code Information
10094143 10094143