Z-2310-2018 Class II Terminated

Recalled by Siemens Medical Solutions USA, Inc — Malvern, PA

FDA device recall Z-2310-2018 was initiated by Siemens Medical Solutions USA, Inc on May 1, 2018 and is designated Class II. Reason for recall: After the Large Display returns from power save mode, it may not show an image, and stay black without showing an error message although X-ray is still possible. The problem does not occur during an… The recall status is terminated (terminated May 16, 2019). Affected quantity: 73.

Recall Details

Product Type
Devices
Report Date
Jul 4, 2018
Initiation Date
May 1, 2018
Termination Date
May 16, 2019
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
73

Product Description

Artis zee ceiling, Material no. 10094137, for angiography and whole body radiographic/fluoroscopic procedures.

Reason for Recall

After the Large Display returns from power save mode, it may not show an image, and stay black without showing an error message although X-ray is still possible. The problem does not occur during an ongoing procedure. If the problem occurs, the system cannot be operated normally. It may be necessary to cancel or restart the treatment or transfer the patient to an alternate or another system.

Distribution Pattern

Worldwide distribution. US nationwide including Puerto Rico, Australia, Austria, Bahrain, Belarus, Belgium, Brazil, Canada, Croatia, Czech Republic, Denmark, Estonia, Finland, France, Germany, Hong Kong, India, Indonesia, Ireland, Israel, Italy, Japan, Luxembourg, Mexico, Myanmar, Netherlands, New Zealand, Norway, China, Paraguay, Philippines, Poland, Portugal, Russia, Saudi Arabia, Singapore, South Africa, South Korea, Spain, Sri Lanka, Sweden, Switzerland, Taiwan, Thailand, Turkey, United Arab Emirates, and United Kingdom.

Code Information

147763 147573 147608 147760 147636 147630 147208 147514 147165 147558 147304 146946 147585 147553 147645 147087 147203 147601 147659 147675 147516 147700 147745 147681 147795 147773 147789 147541 147577 147029 147625 147635 146152 147660 147753 147083 146861 147561 147771 147657 147364 147775 147790 147781 147734 146896 147559 147620 147287 147749 147633 147737 147621 147250 147253 147252 147251 147717 147778 147643 147670 147234 147557 147661 147235 147716 147739 147743 147752 147528 147687 146867 147186

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-2310-2018
openFDA file date
Oct 23, 2026
Loaded by RegDataLab
Oct 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.