Z-2311-2018 Class II Terminated

Recalled by Siemens Medical Solutions USA, Inc — Malvern, PA

Recall Details

Product Type
Devices
Report Date
July 4, 2018
Initiation Date
May 1, 2018
Termination Date
May 16, 2019
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
14

Product Description

Artis zee floor, Material no. 10094135, for angiography and whole body radiographic/fluoroscopic procedures.

Reason for Recall

After the Large Display returns from power save mode, it may not show an image, and stay black without showing an error message although X-ray is still possible. The problem does not occur during an ongoing procedure. If the problem occurs, the system cannot be operated normally. It may be necessary to cancel or restart the treatment or transfer the patient to an alternate or another system.

Distribution Pattern

Worldwide distribution. US nationwide including Puerto Rico, Australia, Austria, Bahrain, Belarus, Belgium, Brazil, Canada, Croatia, Czech Republic, Denmark, Estonia, Finland, France, Germany, Hong Kong, India, Indonesia, Ireland, Israel, Italy, Japan, Luxembourg, Mexico, Myanmar, Netherlands, New Zealand, Norway, China, Paraguay, Philippines, Poland, Portugal, Russia, Saudi Arabia, Singapore, South Africa, South Korea, Spain, Sri Lanka, Sweden, Switzerland, Taiwan, Thailand, Turkey, United Arab Emirates, and United Kingdom.

Code Information

137165 137146 137311 137229 137231 137224 137217 136979 136557 136558 136391 136503 136594 136278