Z-2315-2015 Class II Terminated

Recalled by Tornier, Inc — Bloomington, MN

FDA device recall Z-2315-2015 was initiated by Tornier, Inc on Jun 25, 2015 and is designated Class II. Reason for recall: Recall for the Aequalis IM Nail instrumentation set (Tray Number 9020000) due to several reports that the mast of the targeting jig (Part Number 9020060) is separating from the jig boom. Separation… The recall status is terminated (terminated Aug 31, 2017). Affected quantity: 99.

Recall Details

Product Type
Devices
Report Date
Aug 5, 2015
Initiation Date
Jun 25, 2015
Termination Date
Aug 31, 2017
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
99

Product Description

Aequalis Humeral Nail Targeting Jig, Product Code: 9020060. Orthopedic manual surgical instrument.

Reason for Recall

Recall for the Aequalis IM Nail instrumentation set (Tray Number 9020000) due to several reports that the mast of the targeting jig (Part Number 9020060) is separating from the jig boom. Separation will impact alignment and can affect the ability to fixate Aequalis IM Nail screws.

Distribution Pattern

US: AL, AK, AZ, CA, CO, CT, FL, IL, IN, ME, MD, MI, MO, NC, OH, PA, TX. OUS: CANADA AND FRANCE.

Code Information

Lots: DB0114069, DB0114349, CA46096, CA216052, DB0111264001, DB0112016, DB0113102.