Z-2323-2024 Class II Ongoing
FDA device recall Z-2323-2024 was initiated by Baxter Healthcare Corporation on Jun 3, 2024 and is designated Class II. Reason for recall: During manufacturing, the step to install an Electromagnetic Interference (EMI) Absorber was omitted. The purpose of the EMI absorber is to prevent EMI leakage that might potentially affect other… The recall status is ongoing. Affected quantity: 1 unit.
Recall Details
- Product Type
- Devices
- Report Date
- Jul 17, 2024
- Initiation Date
- Jun 3, 2024
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 1 unit
Product Description
Welch Allyn CP150 Electrocardiograph with Spirometry Option, Product Code: CP150A-1ENB. The Electrocardiograph is one of the tools that clinicians use to evaluate, diagnose, and monitor patient cardiac function.
Reason for Recall
During manufacturing, the step to install an Electromagnetic Interference (EMI) Absorber was omitted. The purpose of the EMI absorber is to prevent EMI leakage that might potentially affect other devices in the vicinity or the device itself may be vulnerable to EMI resulting in device malfunction or unavailability.
Distribution Pattern
Distribution to Illinois.
Code Information
UDI/DI: 00732094177640. Product Code: CP150A-1ENB. Serial Number: 100013892024