Z-2323-2024 Class II Ongoing

Recalled by Baxter Healthcare Corporation — Deerfield, IL

FDA device recall Z-2323-2024 was initiated by Baxter Healthcare Corporation on Jun 3, 2024 and is designated Class II. Reason for recall: During manufacturing, the step to install an Electromagnetic Interference (EMI) Absorber was omitted. The purpose of the EMI absorber is to prevent EMI leakage that might potentially affect other… The recall status is ongoing. Affected quantity: 1 unit.

Recall Details

Product Type
Devices
Report Date
Jul 17, 2024
Initiation Date
Jun 3, 2024
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
1 unit

Product Description

Welch Allyn CP150 Electrocardiograph with Spirometry Option, Product Code: CP150A-1ENB. The Electrocardiograph is one of the tools that clinicians use to evaluate, diagnose, and monitor patient cardiac function.

Reason for Recall

During manufacturing, the step to install an Electromagnetic Interference (EMI) Absorber was omitted. The purpose of the EMI absorber is to prevent EMI leakage that might potentially affect other devices in the vicinity or the device itself may be vulnerable to EMI resulting in device malfunction or unavailability.

Distribution Pattern

Distribution to Illinois.

Code Information

UDI/DI: 00732094177640. Product Code: CP150A-1ENB. Serial Number: 100013892024

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-2323-2024
openFDA file date
Oct 23, 2026
Loaded by RegDataLab
Oct 25, 2026