Z-2345-2015 Class II Terminated

Recalled by The Anspach Effort, Inc. — Palm Beach Gardens, FL

FDA device recall Z-2345-2015 was initiated by The Anspach Effort, Inc. on Jul 1, 2015 and is designated Class II. Reason for recall: Saw head separated from the drive unit or began to come loose. The recall status is terminated (terminated Dec 14, 2016). Affected quantity: 360 handpieces (Currently there are 9 affected units remaining in the US and 17 units OUS all others were reworked).

Recall Details

Product Type
Devices
Report Date
Aug 12, 2015
Initiation Date
Jul 1, 2015
Termination Date
Dec 14, 2016
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
360 handpieces (Currently there are 9 affected units remaining in the US and 17 units OUS all others were reworked)

Product Description

Synthes Battery Power Line II Oscillator Model 530.710, orthopedic surgical power tool.

Reason for Recall

Saw head separated from the drive unit or began to come loose.

Distribution Pattern

Worldwide Distribution - US Distribution to the states of : AR, LA, OH, and WA., and the countries of : China, France, Italy, Russia, South Africa, Switzerland, and United Kingdom.

Code Information

Serial numbers: 100067, 100076, 100371, 100372, 100063, 100065, 100069, 100076, and 100055; Foreign: Item number 530.710 Serial numbers: 100020, 100026, 100039, 100248, 100254, 100257, 100263, 100265, 100269, 100270, 100293, 100296, 100307, 100313, 100315, 100339, and 100340.

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-2345-2015
openFDA file date
Oct 26, 2026
Loaded by RegDataLab
Oct 27, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.