Z-2362-2026 Class II Ongoing

Recalled by The Binding Site Group, Ltd. — Birmingham, N/A

FDA device recall Z-2362-2026 was initiated by The Binding Site Group, Ltd. on Apr 16, 2026 and is designated Class II. Reason for recall: It was reported that plates 2 and 3 were marked as passed , although it appeared that QC had not been performed on these plates, despite samples being measured. The potential clinical hazards include… The recall status is ongoing. Affected quantity: 1.

Recall details

Product Type
Devices
Report Date
Jun 17, 2026
Initiation Date
Apr 16, 2026
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
1

Product description

Brand Name: EXENT Analyser Product Name: EXENT Analyser Model/Catalog Number: IE800.A Software Version: 1.0.20 Product Description: The EXENT Analyser is an automated analyser intended for the qualitative and quantitative in vitro measurement of analytes in human body fluids used in conjunction with the EXENT assays.

Reason for recall

It was reported that plates 2 and 3 were marked as passed , although it appeared that QC had not been performed on these plates, despite samples being measured. The potential clinical hazards include false or missed detection of M-protein (false positive); and falsely elevated or decreased M-protein concentration.

Distribution pattern

US Nationwide distribution.

Code information

Lot Code: Model No; IE800A; EXENT Analyser; UDI; 05051700020886, Serial Number; 101150, Software version; v1.0.20.

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-2362-2026
openFDA file date
Nov 19, 2026
Loaded by RegDataLab
Nov 20, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for The Binding Site Group, Ltd.? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.

Get a free API key

How the connector works

Source: openFDA record for Z-2362-2026. Report a data issue: hello@regdatalab.com.