Z-2375-2018 Class II Terminated

Recalled by Zimmer Biomet, Inc. — Warsaw, IN

FDA device recall Z-2375-2018 was initiated by Zimmer Biomet, Inc. on May 30, 2018 and is designated Class II. Reason for recall: M/L Taper Hip Prosthesis and M/L Taper Hip Prosthesis with Kinectiv Technology manufactured prior to October 2015 are being recalled due to the potential for black residue to be observed within the… The recall status is terminated (terminated May 28, 2020). Affected quantity: 116.

Recall Details

Product Type
Devices
Report Date
Jul 11, 2018
Initiation Date
May 30, 2018
Termination Date
May 28, 2020
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
116

Product Description

Zimmer¿ M/L Taper Hip Prosthesis With Kinective¿ Technology Cementless, Size 9; 00-7713-009-00

Reason for Recall

M/L Taper Hip Prosthesis and M/L Taper Hip Prosthesis with Kinectiv Technology manufactured prior to October 2015 are being recalled due to the potential for black residue to be observed within the polyethylene bag containing the implant. The black residue is a combination of the titanium plasma spray used to create the porous surface of the implant and polyethylene bag containing the implant. The recalling firm states the residue is biocompatible. Risks include delay of surgery while locating a replacement device, and surgical intervention due to wear caused by debris.

Distribution Pattern

Products were distributed solely to Japan.

Code Information

62350575 62647167 62787292 63044896 63044897 63068905 63076554 63120523 63161387 63174034

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-2375-2018
openFDA file date
Oct 23, 2026
Loaded by RegDataLab
Oct 25, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.