Z-2380-2018 Class II Terminated

Recalled by Intuitive Surgical, Inc. — Sunnyvale, CA

FDA device recall Z-2380-2018 was initiated by Intuitive Surgical, Inc. on Apr 23, 2018 and is designated Class II. Reason for recall: Intuitive Surgical, Inc. (ISI) has recently become aware of an issue with the da Vinci Xi Stapler 45 Blue and Green Reloads that may be missing a component of the knife mechanism. Without this… The recall status is terminated (terminated Jul 15, 2020). Affected quantity: 7337.

Recall Details

Product Type
Devices
Report Date
Jul 18, 2018
Initiation Date
Apr 23, 2018
Termination Date
Jul 15, 2020
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
7337

Product Description

da Vinci Xi EndoWrist Stapler 45 Reload Blue, Model 48645B-03, 12 reloads/box.

Reason for Recall

Intuitive Surgical, Inc. (ISI) has recently become aware of an issue with the da Vinci Xi Stapler 45 Blue and Green Reloads that may be missing a component of the knife mechanism. Without this component, the reload knife will not move and cut tissue when the stapler is fired.

Distribution Pattern

U.S., foreign and government accounts. Foreign: Australia, Austria, Belgium, Brazil, Canada, China (Hong Kong), Czech Republic, Denmark, Finland, France, Germany, Ireland, Italy, Japan, Luxembourg, Netherlands, Norway, Qatar, Saudi Arabia, Singapore, South Korea, Spain, Sweden, Switzerland, Taiwan, Turkey, United Kingdom.

Code Information

UDI: 00886874112878. Device expiration: 10/14/2017 02/21/2020

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-2380-2018
openFDA file date
Oct 23, 2026
Loaded by RegDataLab
Oct 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.