Z-2399-2012 Class II Terminated

Recalled by Biomet, Inc. — Warsaw, IN

FDA device recall Z-2399-2012 was initiated by Biomet, Inc. on Aug 14, 2012 and is designated Class II. Reason for recall: Investigation found that the current design of the product allows the press-fit of the knob/shaft interface to be overcome by the impaction force of a mallet, causing the knob to potentially unfasten… The recall status is terminated (terminated Apr 18, 2013). Affected quantity: 62.

Recall Details

Product Type
Devices
Report Date
Sep 26, 2012
Initiation Date
Aug 14, 2012
Termination Date
Apr 18, 2013
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
62

Product Description

JUGGERKNOT IN¿GUIDE PUNCH, 2.9MM. The JuggerKnot 2.9mm In-Guide Punch is intended to be used to prepare a 2.9mm hole into bone in which the JuggerKnot 2.9mm Suture Anchor is to be placed. The JuggerKnot 2.9mm Suture Anchor is intended to be used to attach soft tissue to bone.

Reason for Recall

Investigation found that the current design of the product allows the press-fit of the knob/shaft interface to be overcome by the impaction force of a mallet, causing the knob to potentially unfasten and slide down the shaft.

Distribution Pattern

Nationwide Distribution-including the states of WA, IN, AK, PA, GA, SD, KS, NY, VA, AR, AL, OH, WI, MA, and CA.

Code Information

REF. 912064 LOT 121040, 497540, 839530

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-2399-2012
openFDA file date
Oct 26, 2026
Loaded by RegDataLab
Oct 27, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.