Z-2406-2021 Class II Ongoing
Recall Details
- Product Type
- Devices
- Report Date
- September 8, 2021
- Initiation Date
- July 26, 2021
- Termination Date
- N/A
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- N/A
Product Description
Monaco RTP System, Software Builds 5.40.00, 5.40.01, 5.40.02, and 5.51.10
Reason for Recall
If a couch is removed while a plan is loaded, and then added back to the Studyset when the plan is unloaded, it is possible to calculate dose with the couch unassigned to the beams.
Distribution Pattern
Distribution in US - IA, NJ, NY, TN, PA, TX, and WI OUS distribution in Australia, Bahrain, Belgium, Canada, China, Denmark, France, Germany, Hong Kong, Italy, Japan, Korea, Netherlands, Spain, Sweden, Switzerland, Thailand, Turkey and United Kingdom
Code Information
Software Builds: 5.40.00, 5.40.01, 5.40.02, 5.51.10 UDI: (01)00858164002190(10) 5.40.00 (01)00858164002190(10) 5.40.01 (01)00858164002190(10) 5.40.02 (01)00858164002268(10) 5.51.10