Z-2433-2024 Class II Ongoing

Recalled by B Braun Medical Inc — Breinigsville, PA

FDA device recall Z-2433-2024 was initiated by B Braun Medical Inc on Jun 14, 2024 and is designated Class II. Reason for recall: Perfusor Space Syringe Pump's instructions for use previously identified Monoject syringes as an option for use. However, changes to manufacturing to Cardinal Health Monoject syringes may result in… The recall status is ongoing. Affected quantity: 14949.

Recall Details

Product Type
Devices
Report Date
Jul 31, 2024
Initiation Date
Jun 14, 2024
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
14949

Product Description

Brand Name: Perfusor Space Product Name: Perfusor Space Infusion Pump System Model/Catalog Number: 8713030U Product Description: Perfusor Space Infusion Pump System, Infusion Pump, 1 each Component: N/A

Reason for Recall

Perfusor Space Syringe Pump's instructions for use previously identified Monoject syringes as an option for use. However, changes to manufacturing to Cardinal Health Monoject syringes may result in issues with recognition, compatibility, and pump performance, such as overdose, underdose, delay in therapy, delay in occlusion alarms, and delay in feeding. Cardinal Health Monoject syringes are not compatible with the indicated pumps.

Distribution Pattern

Domestic distribution nationwide. International distribution to Canada.

Code Information

Lot Code: Model No 8713030U; UDI-DI 4046963716745; All Serial Numbers

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-2433-2024
openFDA file date
Oct 23, 2026
Loaded by RegDataLab
Oct 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.