Z-2435-2024 Class II Ongoing

Recalled by B Braun Medical Inc — Breinigsville, PA

Recall Details

Product Type
Devices
Report Date
July 31, 2024
Initiation Date
June 14, 2024
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
1663

Product Description

Brand Name: Perfusor Space Product Name: Perfusor PCA Syringe Pump Model/Catalog Number: 8713080U Product Description: Perfusor¿ Space Syringe Pump, Infusion Pump, 1 each Component: N/A

Reason for Recall

Perfusor Space Syringe Pump's instructions for use previously identified Monoject syringes as an option for use. However, changes to manufacturing to Cardinal Health Monoject syringes may result in issues with recognition, compatibility, and pump performance, such as overdose, underdose, delay in therapy, delay in occlusion alarms, and delay in feeding. Cardinal Health Monoject syringes are not compatible with the indicated pumps.

Distribution Pattern

Domestic distribution nationwide. International distribution to Canada.

Code Information

Lot Code: Model No 8713080U; UDI-DI 4046963915308; All Serial Numbers