FDA device recall Z-2436-2018 was initiated by Siemens Healthcare Diagnostics, Inc. on Mar 21, 2018 and is designated Class II. Reason for recall: The diluent may have an incomplete slit on the septum in the cap of the vial. A diluent cap with a defective septum has the potential to cause erroneously elevated results if a sample is autodiluted… The recall status is terminated (terminated Jun 12, 2020). Affected quantity: 1446.

Recall Details

Product Type
Devices
Report Date
Jul 25, 2018
Initiation Date
Mar 21, 2018
Termination Date
Jun 12, 2020
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
1446

Product Description

Dimension Vista¿ CTNI Sample Diluent, KD692, SMN# 10445205

Reason for Recall

The diluent may have an incomplete slit on the septum in the cap of the vial. A diluent cap with a defective septum has the potential to cause erroneously elevated results if a sample is autodiluted onboard the system, due to a reduced volume of diluent being pipetted from the vial.

Distribution Pattern

The products were distributed to the following US states: AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, and WY. The products were distributed to Guam, Saipan, and Virgin Islands. The products were distributed to the following countries: Australia, Austria, Bahamas, Belgium, Brazil, Canada, Canary Islands, Croatia, Czech Republic, Denmark, France, Germany, Italy, Japan, Lebanon, Netherlands, Poland, Portugal, Qatar, Russian Federation, Slovakia, Slovenia, Saudi Arabia, South Korea, Spain, Switzerland, and United Arab Emirates.

Code Information

lot # 17171BD, UDI # (UDI) 008427680158927BDA3418060110445205

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-2436-2018
openFDA file date
Oct 26, 2026
Loaded by RegDataLab
Oct 27, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.