Z-2438-2018 Class II Terminated

Recalled by Siemens Healthcare Diagnostics, Inc. — Newark, DE

Recall Details

Product Type
Devices
Report Date
July 25, 2018
Initiation Date
March 21, 2018
Termination Date
June 12, 2020
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
1195

Product Description

Dimension Vista¿ MULTI 2 SDIL, KD694, SMN# 10483586

Reason for Recall

The diluent may have an incomplete slit on the septum in the cap of the vial. A diluent cap with a defective septum has the potential to cause erroneously elevated results if a sample is autodiluted onboard the system, due to a reduced volume of diluent being pipetted from the vial.

Distribution Pattern

The products were distributed to the following US states: AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, and WY. The products were distributed to Guam, Saipan, and Virgin Islands. The products were distributed to the following countries: Australia, Austria, Bahamas, Belgium, Brazil, Canada, Canary Islands, Croatia, Czech Republic, Denmark, France, Germany, Italy, Japan, Lebanon, Netherlands, Poland, Portugal, Qatar, Russian Federation, Slovakia, Slovenia, Saudi Arabia, South Korea, Spain, Switzerland, and United Arab Emirates.

Code Information

lot 7CDA79, UDI # (UDI) 008427680324177CDA7918090110483586, exp 2018-09-01; lot # 7DDA70, UDI # (UDI) 008427680324177DDA7018110110483586, exp 2018-11-01