Z-2441-2023 Class II Ongoing

Recalled by Boston Scientific Corporation — Saint Paul, MN

Recall Details

Product Type
Devices
Report Date
August 30, 2023
Initiation Date
July 11, 2023
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
3,856 units

Product Description

EMBLEM S-ICD Pulse Generator Model A209; Subcutaneous Implantable Cardioverter Defibrillator

Reason for Recall

There is a potential for a rare interaction between the EMBLEM S-ICD and LATITUDE communicator, which may cause S-ICD sensing disablement for a 24-hour interval.

Distribution Pattern

Worldwide

Code Information

GTIN: 00802526544101, 00802526548406, 00802526575105, 00802526575112, 00802526575129, 00802526575136, 00802526575143, 00802526575167, 00802526575181, 00802526575204, 00802526575211, 00802526575228, 00802526599002; All EMBLEM S-ICDs enrolled in LATITUDE