Z-1160-2026 Class I Ongoing

Recalled by Boston Scientific Corporation — Marlborough, MA

FDA device recall Z-1160-2026 was initiated by Boston Scientific Corporation on Dec 19, 2025 and is designated Class I. Reason for recall: Increased reports of stent deployment and expansion issues. The most common outcome reasonably foreseeable resulting from difficulty deploying the stent is prolongation of the procedure to exchange… The recall status is ongoing. Affected quantity: 59 units.

Recall details

Product Type
Devices
Report Date
Feb 11, 2026
Initiation Date
Dec 19, 2025
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
59 units

Product description

AXIOS Stent and Electrocautery-Enhanced Delivery System - 8mm x 8mm; Material Number (UPN): M00553690;

Reason for recall

Increased reports of stent deployment and expansion issues. The most common outcome reasonably foreseeable resulting from difficulty deploying the stent is prolongation of the procedure to exchange the device for a new one. The most serious reasonably foreseeable outcome in cases where the first flange is unable to deploy or expand would be a cascade of events resulting in additional intervention, either endoscopic or surgical, to remove the stent and close the puncture site.

Distribution pattern

Worldwide distribution. US Nationwide including Puerto Rico; Algeria, Argentina, Austria, Belgium, Brazil, Canada, CHILE, Colombia, Costa Rica, Croatia, Czech Republic, Denmark, Finland, France, Germany, Greece, Hungary, India, Indonesia, Iran, Ireland, Israel, Italy, JAPAN, Jordan, Kuwait, Lebanon, Luxembourg, Malaysia, Mexico, Morocco, Netherlands, Norway, Oman, Pakistan, Palestine, Panama, Paraguay, PERU, Poland, Portugal, Romania, Saudi Arabia, Serbia, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, UAE, and United Kingdom.

Code information

Material Number (UPN): M00553690; UDI-DI: 191506008093; Lot Numbers: 35960775, 35960777, 35962525, 36100561, 36101748, 36111316, 36154532, 36154533, 36241501, 36251362, 36253016, 36293839, 36301625, 36303192, 36420219, 36425888, 36449561, 36458830, 36461449, 36478580, 36492661, 36495673, 36573597, 36573598, 36573599, 36574152, 36576474, 36584299, 36604665, 36731807, 36731808, 36737997, 36743059, 36743200, 36904231, 36904232, 36904233, 36906587, 36907000, 36907001, 36907032, 37217612, 37217613, 37217614, 37440330, 37495327, 37511242, 37602825, 37602826, 37689946, 37694053, 37694054, 37764851, 37764852, 37995048, 37995049, 37995050, 38025042, 38025043,

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1160-2026
openFDA file date
Nov 25, 2026
Loaded by RegDataLab
Nov 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Boston Scientific Corporation? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for Boston Scientific Corporation: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about Boston Scientific Corporation →

Source: openFDA record for Z-1160-2026. Report a data issue: hello@regdatalab.com.