Z-2449-2015 Class II Terminated

Recalled by Carestream Health Inc — Rochester, NY

FDA device recall Z-2449-2015 was initiated by Carestream Health Inc on Aug 7, 2015 and is designated Class II. Reason for recall: Software defect that impacts image alignment when using the automatic stitching option for long length images. The defect could cause stitching inconsistencies which may go undetected on a radiograph. The recall status is terminated (terminated Jul 6, 2016). Affected quantity: Domestic: 76 units, Foreign: 160 units.

Recall details

Product Type
Devices
Report Date
Sep 2, 2015
Initiation Date
Aug 7, 2015
Termination Date
Jul 6, 2016
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
Domestic: 76 units, Foreign: 160 units

Product description

CARESTREAM DRX-EVOLUTION X-Ray System Product Usage: The device is a permanently installed diagnostic X-ray system for general radiographic x-ray imaging including tomography. The tomography feature is not to be used for imaging pediatric patients.

Reason for recall

Software defect that impacts image alignment when using the automatic stitching option for long length images. The defect could cause stitching inconsistencies which may go undetected on a radiograph.

Distribution pattern

Worldwide Distribution - US Nationwide and in the countries: Australia, Canada, Chile, China, India, Korea, Malaysia, New Zealand, Singapore, Switzerland, Taiwan and Thailand.

Code information

Software versions 5.6 and greater and 5.7 (current releases); Serial Numbers: 5029, 5569, 4067, 4754, 4131, 4100, 4096, 4076, 4569, 4570, 4588, 4670, 4767, 4879, 5266, 5393, 5408, 5442, 5572, 5573, 5555, 5556, 5563, 4146, 4679, 4702, 4703, 4704, 4801, 5090, 5154, 4787, 4069, 4074, 4344, 4085, 4058, 4084, 5470, 5158, 4086, 4099, 4073, 4665, 5241, 5303, 5304, 4639, 5350, 5352, 5429, 5508, 4696, 4763, 5182, 4291, 5037, 5554, 4192, 4117, 4172, 4207, 4225, 4239, 4278, 4301, 4310, 4328, 4333, 4338, 4366, 4384, 4421, 4427, 4442, 4443, 4447, 4465, 4499, 4523, 4534, 4538, 4539, 4692, 4822, 4861, 4875, 5128, 5131, 5174, 5553, 5579, 5580, 5597, 5628, 4320, 5309, 5042, 4807, 4808, 4809, 4810, 5493, 4266, 4711, 5327, 4125, 4143, 4097, 4156, 4282, 4287, 4289, 4258, 4742, 4964, 4784, 5050, 5551, 4233, 4644, 4082, 4113, 4196, 4683, 4309, 4315, 4064, 4256, 4987, 4624, 4585, 4882, 5362, 4004, 5377, 5418, 4300, 4308, 4627, 4638, 5138, 5251, 5360, 4947, 5620, 4985, 5403, 5495, 5560, 5544, 5545, 5621, 5622, 5451, 5282, 5645, 4724, 4298, 4269, 4756, 4418, 5273, 5178, 5179, 5443, 5197, 5389, 5584, 5473, 4899, 5334, 5431, 4990 --- NOTE: Additional serial numbers at foreign consignees to be determined upon reporting part usage.

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-2449-2015
openFDA file date
Nov 19, 2026
Loaded by RegDataLab
Nov 20, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.