Z-1052-2017 Class II Terminated

Recalled by Carestream Health Inc — Rochester, NY

Recall Details

Product Type
Devices
Report Date
January 25, 2017
Initiation Date
November 21, 2016
Termination Date
March 7, 2017
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
US: 14 units; Foreign: 16 units

Product Description

Carestream Touch Prime, Catalog # 1738830, and Touch Prime XE, Catalog # 1738822, Ultrasound System Diagnostic ultrasound imaging or fluid flow analysis of the human body

Reason for Recall

Software error; Carestream Health Inc, received a complaint stating that when a user accidentally obtains a measurement value of 0 and corrects the value in the report, the resulting measurement unit is not displayed, i.e., centimeters or millimeters. As such, the user expects that the measurement is taken calculated in centimeters, consistent with other values in the report. In actuality, the measurement is taken in millimeters. When this updated measurement is used in an average calculation, the result appears incorrect as two measurements are interpreted as centimeters while the user corrected value is interpreted as millimeters. If the user selects a Calc Result display as Min or Max, the values are also interpreted as millimeters when centimeters were expected.

Distribution Pattern

Worldwide Distribution - US including SC, GA, TX, OH, IA, NY and foreign distribution to Switzerland and Malaysia

Code Information

Serial Numbers: US units: 106, 112, 114, 116, 1021, 1037, 1038, 1044, 1057, 1067, 1072, 1074, 1079, 1090; OUS units: 1005, 1006, 1009, 1011, 1015, 1017, 1018, 1020, 1031, 1041, 1042, 1043, 1045, 1081, 1086, 1089.