Z-2465-2015 Class II Terminated

Recalled by Tornier, Inc — Bloomington, MN

FDA device recall Z-2465-2015 was initiated by Tornier, Inc on Aug 6, 2015 and is designated Class II. Reason for recall: Tornier has initiated a voluntary recall for the Salto Talaris Tibial Guide (part number MJU333) due to a complaint that occurred during physician training, that the guide cannot be assembled to the… The recall status is terminated (terminated Apr 8, 2016). Affected quantity: 5.

Recall Details

Product Type
Devices
Report Date
Sep 2, 2015
Initiation Date
Aug 6, 2015
Termination Date
Apr 8, 2016
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
5

Product Description

Salto Talaris Tibial Guide, Part # MJU333 ,Tray/Kit Code YKAL11, Non sterile. Orthopedic Manual Surgical Instrument

Reason for Recall

Tornier has initiated a voluntary recall for the Salto Talaris Tibial Guide (part number MJU333) due to a complaint that occurred during physician training, that the guide cannot be assembled to the Salto Talaris Talar Pin Setting guide. This issue will cause the impossibility the time of putting the Talar pin in the guide.

Distribution Pattern

Distributed to CA, ME, NC, PA, and TX.

Code Information

Lot #14F005

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-2465-2015
openFDA file date
Oct 23, 2026
Loaded by RegDataLab
Oct 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.