Z-2471-2018 Class II Ongoing
Recall Details
- Product Type
- Devices
- Report Date
- July 25, 2018
- Initiation Date
- April 23, 2018
- Termination Date
- N/A
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 55
Product Description
t:slim G4 Insulin Pump with Dexcom G4 Platinum CGM Software version: 4.3.4.3, Firmware version: 004722, Firmware, Fuel Gauge, Binary Rev B.
Reason for Recall
The fuel gauge, the component that reads and reports the battery parameters to the pump, could provide inaccurate readings, which present to the user in one of two ways: A) by triggering a Malfunction 4 Alarm, or B) by triggering a succession of notifications prior to the pump shutting off, including Alert 2 (Battery Low), Alert 3 (Battery Very Low) and Alarm 12 (Battery Very Low). The alarms in both scenarios notify the user that the pump has stopped delivering insulin.
Distribution Pattern
U.S. Consignees: IN, AZ, UT, TX, NY, NJ, MD, MN, IL, CO, AL, WI, CA, PA, OH, OR, MS, FL, NM, ID, NV, KY, VA,
Code Information
Serial Numbers: 483731, 483930, 483992, 484013, 492055, 492368, 501157, 501177, 501178, 501187, 501191, 501200, 501202, 501205, 501207, 501268, 501285, 501364, 501368, 501370, 501376, 501385, 501386, 501387, 501432, 501435, 501440, 501443, 501444, 501445, 501466, 501479, 501488, 501490, 501491, 501495, 501503, 501508, 501509, 501519, 501522, 501523, 501526, 501535, 501560, 504522, 504531, 504535, 504540, 504753, 504972, 504990, 505025, 90472671, 90488584