Z-1609-2024 Class I Ongoing

Recalled by Tandem Diabetes Care, Inc. — San Diego, CA

FDA device recall Z-1609-2024 was initiated by Tandem Diabetes Care, Inc. on Mar 5, 2024 and is designated Class I. Reason for recall: During normal use, the mobile app version 2.7 may crash and be automatically relaunched by the iOS operating system. This cycle intermittently repeats, which leads to excessive Bluetooth communication… The recall status is ongoing. Affected quantity: 85,863 applications.

Recall details

Product Type
Devices
Report Date
May 15, 2024
Initiation Date
Mar 5, 2024
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
85,863 applications

Product description

t:connect mobile app used in conjunction with t:slim X2 insulin pump with Control-IQ technology

Reason for recall

During normal use, the mobile app version 2.7 may crash and be automatically relaunched by the iOS operating system. This cycle intermittently repeats, which leads to excessive Bluetooth communication that may result in pump battery drain and may lead to the pump shutting down sooner than typically expected. Pump shutdown will cause insulin delivery to suspend, which could lead to an under-delivery of insulin and may result in hyperglycemia, including severe hyperglycemia.

Distribution pattern

US nationwide distribution including Puerto Rico.

Code information

version 2.7 or later /UDI: 00850006613274

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1609-2024
openFDA file date
Nov 25, 2026
Loaded by RegDataLab
Nov 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Tandem Diabetes Care, Inc.? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.

Get a free API key

How the connector works

Source: openFDA record for Z-1609-2024. Report a data issue: hello@regdatalab.com.