Z-0437-2027 Class II Ongoing

Recalled by Tandem Diabetes Care, Inc. — San Diego, CA

FDA device recall Z-0437-2027 was initiated by Tandem Diabetes Care, Inc. on Oct 13, 2026 and is designated Class II. Reason for recall: A software defect in the Tandem Mobi mobile application for Android (versions 3.5.4, 3.5.2.2, and 3.5.2.1) in which the immediate ( Now ) portion of an Extended Bolus is incorrectly reduced to 0 units… The recall status is ongoing. Affected quantity: 4086 users.

Recall details

Product Type
Devices
Report Date
Nov 18, 2026
Initiation Date
Oct 13, 2026
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
4086 users

Product description

Tandem Mobi Insulin Pump with Interoperable Technology (with Android Tandem Mobi Mobile app) Model/Catalog Number: 1015472

Reason for recall

A software defect in the Tandem Mobi mobile application for Android (versions 3.5.4, 3.5.2.2, and 3.5.2.1) in which the immediate ( Now ) portion of an Extended Bolus is incorrectly reduced to 0 units when the entered glucose value equals the user s active target glucose value; the deferred ( Later ) portion is delivered, but the total insulin delivered may be less than intended, and the application does not provide a distinct alert that the immediate portion was omitted, potentially resulting in unrecognized under-delivery of insulin and hyperglycemia.

Distribution pattern

Worldwide - US Nationwide and the country of Czech Republic.

Code information

Lot Code: UDI-DI: 00389152547244 Software versions: 3.5.4, 3.5.2.2, and 3.5.2.1

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0437-2027
openFDA file date
Nov 24, 2026
Loaded by RegDataLab
Nov 25, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Tandem Diabetes Care, Inc.? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.

Get a free API key

How the connector works