Z-3330-2026 Class II Ongoing

Recalled by Tandem Diabetes Care, Inc. — San Diego, CA

FDA device recall Z-3330-2026 was initiated by Tandem Diabetes Care, Inc. on Jul 16, 2026 and is designated Class II. Reason for recall: A beta (early-release) version of the iOS application, version 2.13 (TestFlight) had a software issue preventing successful pairing of the t:slim application with Abbott FreeStyle Libre 3 Plus (FSL3P)… The recall status is ongoing. Affected quantity: 177.

Recall details

Product Type
Devices
Report Date
Oct 7, 2026
Initiation Date
Jul 16, 2026
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
177

Product description

Brand Name: Tandem t:slim mobile application Product Name: t:slim Mobile Application, iOS Model/Catalog Number: iOS software version 2.13 Software Version: iOS software version 2.13 UDI-DI: 00389152470740 Component: No

Reason for recall

A beta (early-release) version of the iOS application, version 2.13 (TestFlight) had a software issue preventing successful pairing of the t:slim application with Abbott FreeStyle Libre 3 Plus (FSL3P) continuous glucose monitoring (CGM) sensors, resulting in the insulin pump being unable to establish a near-field communication (NFC) connection with the CGM and requiring FSL3P users to use their t:slim insulin pumps in manual mode. FSL3P users are unable to access their CGM's glucose values from within the t:slim app and cannot use their pumps in automated control (closed-loop) mode, which could lead to users experiencing hypoglycemia or hyperglycemia.

Distribution pattern

US Nationwide distribution.

Code information

Lot Code: UDI-DI: 00389152470641 Software Version: 2.13 (iOS only)

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-3330-2026
openFDA file date
Nov 10, 2026
Loaded by RegDataLab
Nov 11, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Source: openFDA record for Z-3330-2026. Report a data issue: hello@regdatalab.com.