Z-2484-2018 Class II Terminated

Recalled by Stryker Sustainability Solutions — Tempe, AZ

Recall Details

Product Type
Devices
Report Date
July 25, 2018
Initiation Date
May 8, 2018
Termination Date
July 29, 2019
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
180 units

Product Description

Reprocessed HARMONIC Ace¿ +7, 5mm Diameter Shears with Advanced Hemostasis (HARH), Item Number HARH36 Product Usage: The Reprocessed HARMONIC ACE¿+7, 5 mm Diameter Shears with Advanced Hemostasis are indicated for soft tissue incisions when bleeding control and minimal thermal injury are desired. The instruments can be used as an adjunct to or substitute for electrosurgery, lasers and steel scalpels in general, plastic, pediatric, gynecologic, urologic, thoracic, exposure to orthopedic structures (such as spine and joint space), sealing and transection of lymphatic vessels, and other open and endoscopic procedures. The instruments allow for the coagulation of vessels up to and including 7 mm in diameter, using the Advanced Hemostasis hand controls button.

Reason for Recall

Stryker Sustainability Solutions has received an increase in reports indicating that the affected Reprocessed HARH36 devices may display an error code (No Instrument Uses Remaining) upon initial connection to the generator. When this error code is present, the device is not able to be used.

Distribution Pattern

US Nationwide Distribution in the states of PA, FL, MD, MA, CA, NY, NC, TX, MT, AL, LA, TN

Code Information

Lot number 7656512