Z-1315-2023 Class II Ongoing

Recalled by Stryker Sustainability Solutions — Tempe, AZ

FDA device recall Z-1315-2023 was initiated by Stryker Sustainability Solutions on Mar 1, 2023 and is designated Class II. Reason for recall: Increased incidence of disposable tourniquet cuffs "not able to achieve desired pressure" and "not holding pressure" due to 1) Cuff fabric tearing during use, and 2) Pinched or unseated O-ring at the… The recall status is ongoing. Affected quantity: 127,894.

Recall details

Product Type
Devices
Report Date
Apr 5, 2023
Initiation Date
Mar 1, 2023
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
127,894

Product description

Color Cuff Single Port, Single Bladder 30 (Royal Blue) Quick Connect, REF: 5921-030-135; Color Cuff Dual Port, Single Bladder 30 (Royal Blue) Quick Connect, REF: 5921-030-235; Color Cuff Single Port, Single Bladder 34 (Purple), Quick Connect, REF: 5921-034-135; Color Cuff Dual Port, Single Bladder 34 (Purple), Quick Connect, REF: 5921-034-235; Color Cuff Single Port, Single Bladder 18 (Red), Quick Connect, REF: 5921-218-135; Color Cuff Dual Port, Single Bladder 18 (Red), Quick Connect, REF: 5921-218-235;

Reason for recall

Increased incidence of disposable tourniquet cuffs "not able to achieve desired pressure" and "not holding pressure" due to 1) Cuff fabric tearing during use, and 2) Pinched or unseated O-ring at the connection between the cuff, disposable fill line, and/or cuff hosing, which may lead to loss of vascular occlusion, which may lead to operative site blood loss, hemorrhage, or other complications.

Distribution pattern

US: WI, MD, NY, PA, WA, CO, GA, FL, OH, NC, TX, NJ, OR, MT, CA, RI, SC, NE, MI, LA, OK, SD, AL, VA, MS, WV, AZ, KY, IL, AR, ME, ND, MO, KS, HI, NH, AK, TN, IA, MN, IN, NV, MA, ID, DE, NM, WY, DC, VT, CT

Code information

REF/UI-DI: 5921-030-135/885825016142, 5921-030-235/885825016166, 5921-034-135/885825016173, 5921-034-235/885825016197, 5921-218-135/885825016500, 5921-218-235/885825016296 Lots: 0000202775, 0000209215, 0000202137, 0000201380, 0000200708, 13872322, 13453772, 0000198235, 0000202066, 0000209420, 0000206425, 0000192978, 0000202577, 0000192465, 13463347, 0000204578, 0000193181, 0000198446, 0000208949, 0000195337, 0000206263, 0000193179, 13486613, 0000202777, 0000195071, 0000204159, 0000195912, 0000200905, 0000206442, 0000197869, 13594486, 0000192979, 0000200706, 0000195068, 0000201379, 0000199502, 0000199434, 0000188863, 13493683, 0000193177, 0000202065, 0000200712, 0000200242, 0000202138, 13582864, 0000195075, 13570832, 0000199808, 0000202075, 0000195340, 0000200254, 0000202932, 13774680, 13851472, 13634704, 0000199819, 0000204319, 0000204160, 0000201378, 0000200227, 0000203311, 0000206583, 13235119, 0000194243, 0000196364, 0000193841, 0000200270, 13828088, 0000196363, 13392486, 13280380, 0000192981, 0000198461, 0000196361, 0000200244, 0000203310, 0000200229, 13398036, 13352572, 0000194232, 0000200253, 0000198245, 0000202776, 0000197373, 0000187746, 13765829, 13449838, 0000200256, 0000202145, 0000198246, 0000198236, 0000204574, 0000202143, 0000194242, 13583025, 0000202074, 0000203315, 0000195073, 0000193176, 0000212764, 0000196379, 13811283, 13605685, 0000187310, 0000208444, 0000204366, 0000197867, 13865381, 13820702, 0000176566, 13663274, 0000196365, 0000210787, 0000206440, 0000193391, 0000201391, 0000193412, 0000177131, 13883084, 0000198460, 0000208345, 0000206443, 0000192466, 0000210962, 0000200230, 0000204570, 13339407, 0000204314, 0000211530, 0000209200, 0000196359, 0000210005, 0000201840, 0000205922, 0000192980, 0000205238, 0000195335, 0000206518, 0000195067, 0000203313, 0000193393, 0000206254, 0000208440, 0000202686, 0000198244, 0000203314, 0000195069, 0000208407, 0000206432, 13632837, 0000210465, 0000206505, 0000172873, 0000195338, 0000199805, 0000208321, 0000195343, 0000205241, 0000200707, 0000199807, 0000195076, 0000197375, 0000203312, 13870728, 0000202139, 0000206511, 0000193392, 13824462, 0000195070, 0000199816, 0000208410, 13871893, 0000199503, 0000206586, 13272167, 13399919, 0000195072, 0000199817, 0000192977, 0000197872, 0000186596, 0000208322, 13573489, 0000209992, 0000202684, 0000211154, 0000201837, 0000210310, 0000197870, 0000208606, 0000204571, 13883045, 0000197376, 0000194228, 0000204195, 0000198237, 0000212878, 0000209416, 0000210963, 0000210601, 0000200711, 0000194229, 0000204372, 0000196362, 13807840, 0000210008, 0000200228, 0000172314, 0000200257, 0000194230, 0000205930, 0000193839, 0000193178, 0000210313, 0000198459, 13814050, 0000202144, 0000193413, 0000195336, 13765599, 0000208951, 0000195078, 0000194227, 0000193838, 0000193840, 0000202778

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1315-2023
openFDA file date
Nov 27, 2026
Loaded by RegDataLab
Nov 28, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Stryker Sustainability Solutions? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for Stryker Sustainability Solutions: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about Stryker Sustainability Solutions →

Source: openFDA record for Z-1315-2023. Report a data issue: hello@regdatalab.com.