Z-2489-2025 Class II Ongoing

Recalled by RAYSEARCH LABORATORIES AB — Stockholm, N/A

Recall Details

Product Type
Devices
Report Date
September 10, 2025
Initiation Date
August 8, 2025
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
26

Product Description

RayStation with the following product descriptions: 1. RayStation 9B, Software Version: 9.1.0.933. 2. RayStation 9B Service Pack 1, Software Version: 9.2.0.483. 3. RayStation 10A, Software Version: 10.0.0.1154. 4. RayStation 10A Service Pack 1, Software Version: 10.0.1.52. 5. RayStation 10A Service Pack 2, Software Version: 10.0.2.10. Product Description: Radiation Therapy Treatment Planning System.

Reason for Recall

DICOM SOP Instance UID and Series Instance UID from RayGateway, i.e. the interface to Accuray's iDMS, are not guaranteed to be unique.

Distribution Pattern

Worldwide - US Nationwide distribution in the states of AL, CA, FL, GA, IL, IA, KY, LA, MD, MN, MS, NY, NC, OH, OK, SC, WA, WI and the countries of Canada, China, Colombia, France, Germany, Hong Kong S.A.R., China, Italy, Japan, Netherlands, Poland, Spain, Sweden, Switzerland, ENG, Thailand, Turkey, UK, Ukraine.

Code Information

1. Software Version: 9.1.0.933, UDI: 0735000201026620191220. 2. Software Version: 9.2.0.483, UDI: 0735000201029720200310. 3. Software Version: 10.0.0.1154, UDI: 0735000201030320200526. 4. Software Version: 10.0.1.52, UDI: 0735000201036520200526. 5. Software Version: 10.0.2.10, UDI: 0735000201065520220608. Expiration Date: 2025-06-16