Z-2517-2024 Class II Ongoing

Recalled by American Contract Systems, Inc. — Kansas City, MO

FDA device recall Z-2517-2024 was initiated by American Contract Systems, Inc. on May 6, 2024 and is designated Class II. Reason for recall: Product released with insufficient aeration time impacting the kits sterilized on the same cycle or in the same aeration chamber, which may have been compromised in sterility The recall status is ongoing. Affected quantity: 90.

Recall Details

Product Type
Devices
Report Date
Aug 14, 2024
Initiation Date
May 6, 2024
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
90

Product Description

American Contract Systems Hand Pack convenience kit

Reason for Recall

Product released with insufficient aeration time impacting the kits sterilized on the same cycle or in the same aeration chamber, which may have been compromised in sterility

Distribution Pattern

US Nationwide distribution in the states of CA, KS, ME, MN, MO, NE, and SD.

Code Information

UDI-DI 00191072203717; Lot 982241;Exp. Date 1/18/2025.

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-2517-2024
openFDA file date
Oct 23, 2026
Loaded by RegDataLab
Oct 24, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.