Z-2524-2018 Class II Terminated

Recalled by Siemens Medical Solutions USA, Inc — Malvern, PA

FDA device recall Z-2524-2018 was initiated by Siemens Medical Solutions USA, Inc on May 30, 2018 and is designated Class II. Reason for recall: During manual LAO/RAO rotation at maximum speed or a combination of LAO/RAO and Cran/Caud rotation at maximum speed, the possibility exists for the safety monitoring within the system control to… The recall status is terminated (terminated Feb 14, 2020). Affected quantity: 40 in the United States.

Recall Details

Product Type
Devices
Report Date
Aug 1, 2018
Initiation Date
May 30, 2018
Termination Date
Feb 14, 2020
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
40 in the United States

Product Description

ARTIS Pheno, Fluoroscopic X-Ray system, Model Number 10849000

Reason for Recall

During manual LAO/RAO rotation at maximum speed or a combination of LAO/RAO and Cran/Caud rotation at maximum speed, the possibility exists for the safety monitoring within the system control to become activated. As a result, the C-arm stops the rotation and can drop up to 40 mm (1.6 ).

Distribution Pattern

The devices were distributed to the following US states: AL, CA, CO, FL, IL, KY, MA, MI, MN, MS, ME, NJ, NY, OH, TN, TX, UT, and VA.

Code Information

164033, 164071, 164034, 164095, 164057, 164054, 164055, 164080, 164072, 164088, 164048, 164094, 164068, 164019, 164013, 164073, 164025, 164026, 164027, 164060, 164083, 164021, 164066, 164075, 164074, 164084, 164086, 164098, 164028, 164053, 164063, 164064, 164024, 164049, 164018, 164022, 164061, 164100, 164097, 164011

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-2524-2018
openFDA file date
Oct 23, 2026
Loaded by RegDataLab
Oct 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.