Z-2524-2024 Class II Ongoing

Recalled by American Contract Systems, Inc. — Kansas City, MO

Recall Details

Product Type
Devices
Report Date
August 14, 2024
Initiation Date
May 6, 2024
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
3

Product Description

American Contract Systems Spine Pack convenience kit

Reason for Recall

Product released with insufficient aeration time impacting the kits sterilized on the same cycle or in the same aeration chamber

Distribution Pattern

US Nationwide distribution in the states of CA, KS, ME, MN, MO, NE, and SD.

Code Information

UDI-DI 00191072202499; Lot 980241; Exp. Date 1/10/2025.