Z-2525-2018 Class II Terminated

Recalled by Siemens Medical Solutions USA, Inc — Malvern, PA

Recall Details

Product Type
Devices
Report Date
August 1, 2018
Initiation Date
May 25, 2018
Termination Date
May 16, 2019
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
6 units US

Product Description

AXIOM Artis with Laird Cooling Unit for SSFD: Material # 7555118. Artis is an angiography system developed for single and biplane diagnostic imaging and interventional procedures.

Reason for Recall

Due to a defective sealing, coolant may inflow into the electrical parts of the system cabinet, which could cause a short circuit and cause the system to shut down suddenly during an ongoing procedure

Distribution Pattern

HI , IL, MI, NC, NY, VA, WI

Code Information

Laird Cooling Unit for SSFD: Material # 7555118.