Z-2545-2014 Class II Terminated

Recalled by Zimmer, Inc. — Warsaw, IN

FDA device recall Z-2545-2014 was initiated by Zimmer, Inc. on Jul 30, 2014 and is designated Class II. Reason for recall: Internal review found that five (5) PFJ implant forgings were identified as have forged laps during receiving inspection process for raw material. These five forgings were not segregated and were… The recall status is terminated (terminated Jan 20, 2015). Affected quantity: 15 units.

Recall Details

Product Type
Devices
Report Date
Sep 3, 2014
Initiation Date
Jul 30, 2014
Termination Date
Jan 20, 2015
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
15 units

Product Description

Zimmer Gender Solution Patello-Femoral Joint (PFJ) System Patello-Femoral Trochlea Component, Precoat, Size 4, Left. Knee prosthesis.

Reason for Recall

Internal review found that five (5) PFJ implant forgings were identified as have forged laps during receiving inspection process for raw material. These five forgings were not segregated and were subsequently issued to final manufacturing under Part 00-5926-014-01; lot 62455332. A forged lap is a defect that forms when metal folds over itself during forging.

Distribution Pattern

Worldwide Distribution - US (nationwide) in the states of HI, CA, GA, MI, and IL and the countries of TAIWAN, BELGIUM, AUSTRALIA, FRANCE, JAPAN, and UK.

Code Information

00-5926-014-01; Lot 62455332

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-2545-2014
openFDA file date
Oct 23, 2026
Loaded by RegDataLab
Oct 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.