Z-2546-2014 Class II Terminated

Recalled by Zimmer, Inc. — Warsaw, IN

FDA device recall Z-2546-2014 was initiated by Zimmer, Inc. on Jul 22, 2014 and is designated Class II. Reason for recall: Zimmer is recalling sterile-packaged implants and instruments packaged in six different package configurations due to packaging design verification test failures. Specifically, multiple test samples… The recall status is terminated (terminated Sep 15, 2015). Affected quantity: 972,125 units.

Recall Details

Product Type
Devices
Report Date
Sep 3, 2014
Initiation Date
Jul 22, 2014
Termination Date
Sep 15, 2015
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
972,125 units

Product Description

Knee Instrument Fixation Screws: 48MM HEADED SCREW 48MM HEADLESS SCREW MIS HEADED SCREW, 48MM MIS HEADLESS SCREW, 48MM

Reason for Recall

Zimmer is recalling sterile-packaged implants and instruments packaged in six different package configurations due to packaging design verification test failures. Specifically, multiple test samples from each of the six configurations failed simulated distribution and shipping testing when the devices either breached inner or outer package trays or seal integrity was lost.

Distribution Pattern

Worldwide Distribution.

Code Information

Item Numbers 00579104100, 00579104200, 00598304048, 00598304148 Exp Date Prior to 2023-11

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-2546-2014
openFDA file date
Oct 23, 2026
Loaded by RegDataLab
Oct 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.