Z-2549-2019 Class II Terminated

Recalled by Philips North America LLC — Bothell, WA

FDA device recall Z-2549-2019 was initiated by Philips North America LLC on Aug 8, 2018 and is designated Class II. Reason for recall: The recall is being initiated due to incorrect age range labeling (8+ yrs) on the foil packaging for infant/child pad cartridges. The infant/child pad cartridge is designed for use with the… The recall status is terminated (terminated Apr 29, 2021). Affected quantity: 181,894 Infant/Child SMART Pads Pads.

Recall Details

Product Type
Devices
Report Date
Oct 2, 2019
Initiation Date
Aug 8, 2018
Termination Date
Apr 29, 2021
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
181,894 Infant/Child SMART Pads Pads

Product Description

Philips HeartStart Infant/Child SMART Pads Cartridge, Model # M5072A, UDI:00884838023758

Reason for Recall

The recall is being initiated due to incorrect age range labeling (8+ yrs) on the foil packaging for infant/child pad cartridges. The infant/child pad cartridge is designed for use with the defibrillator An Infant/Child pad cartridge should not be used on an adult because it delivers lower energy (50 J vs 150 J) than the Adult pads cartridge and may therefore not be effective on an adult.

Distribution Pattern

US: AK, AL,AR,AZ,CA,CO,CT,DE,FL,GA,HI,IA,ID,IL,IN,KS,KY,LA,MA,MD,ME,MI,MN,MO,MS,MT,NC,ND,NE,NH,NJ,NM,NV,NY,OH,OK,OR,PA,PR,RI,SC,SD,TN,TX,UT,VA,VT,WA,WI,WV, WY, Argentina, Aruba, Australia, Austria, Belgium, Bolivia, Cambodia, Canada, Chile, China, Colombia, Costa Rica, Cyprus, Denmark, Dominican Republic, Egypt, El Salvador, France, French Guiana, Germany, Greece, Guadeloupe, Hong Kong, India, Indonesia, Ireland, Israel, Italy, Japan, Korea, Republic of, Malaysia, Malta, Martinique, Mauritius, Mexico, Netherlands, Netherlands Antilles, New Caledonia, New Zealand, Norway, Panama, Paraguay, Peru, Philippines, Poland, Puerto Rico, R¿union, Singapore, South Africa, Spain, Sweden, Switzerland, Taiwan, Trinidad and Tobago, United Arab Emirates, United Kingdom, Uruguay, Viet Nam,

Code Information

Model# M5072A; Lots #s Y070115-04 - Y011118-07;

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-2549-2019
openFDA file date
Oct 23, 2026
Loaded by RegDataLab
Oct 24, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.