Z-2602-2016 Class II Terminated

Recalled by Medtronic Neuromodulation — Minneapolis, MN

FDA device recall Z-2602-2016 was initiated by Medtronic Neuromodulation on Jun 10, 2016 and is designated Class II. Reason for recall: Medtronic has voluntarily decided to retrieve unused Model 3387S-40 and 3389S-40 DBS leads from three manufacturing lot numbers. During the manufacturing process of a DBS lead component, Medtronic… The recall status is terminated (terminated Oct 3, 2016). Affected quantity: 70.

Recall Details

Product Type
Devices
Report Date
Aug 31, 2016
Initiation Date
Jun 10, 2016
Termination Date
Oct 3, 2016
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
70

Product Description

Medtronic, Lead Kit for DBS Stimulation. Model Numbers 3387S-40 (p/n 3387S0007V) & 3389S-40 (p/n 3389S0008V).

Reason for Recall

Medtronic has voluntarily decided to retrieve unused Model 3387S-40 and 3389S-40 DBS leads from three manufacturing lot numbers. During the manufacturing process of a DBS lead component, Medtronic identified the potential for lead insulation damage.

Distribution Pattern

US: AR, CA, FL, IL MD, MO, NE, NY, OH, TN WI.

Code Information

lots: VA15GPJ, VA15K3N, VA15K7K