Z-2603-2018 Class II Terminated
FDA device recall Z-2603-2018 was initiated by Medtronic Sofamor Danek USA Inc on Jul 9, 2018 and is designated Class II. Reason for recall: This lot has incorrectly been laser marked both on the screw head itself as well as the data carrier tag attached to the screw. The description on face two of the screw head and on the data carrier… The recall status is terminated (terminated May 7, 2021). Affected quantity: 47 units.
Recall Details
- Product Type
- Devices
- Report Date
- Aug 15, 2018
- Initiation Date
- Jul 9, 2018
- Termination Date
- May 7, 2021
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 47 units
Product Description
Medtronic CD HORIZON Spinal System is intended to help provide immobilization and stabilization of spinal segments as an adjunct to fusion of the thoracic, lumbar, or sacral spine.MULTI-AXIAL SCREW FOR 5.5/6.0MM RODS, REF 55840005045
Reason for Recall
This lot has incorrectly been laser marked both on the screw head itself as well as the data carrier tag attached to the screw. The description on face two of the screw head and on the data carrier tag both incorrectly indicate a 5.5 mm diameter screw when the actual screw diameter aligns with 5.0 diameter screw size requirements.
Distribution Pattern
US
Code Information
lot 0648247W