Z-2603-2018 Class II Terminated

Recalled by Medtronic Sofamor Danek USA Inc — Memphis, TN

Recall Details

Product Type
Devices
Report Date
August 15, 2018
Initiation Date
July 9, 2018
Termination Date
May 7, 2021
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
47 units

Product Description

Medtronic CD HORIZON Spinal System is intended to help provide immobilization and stabilization of spinal segments as an adjunct to fusion of the thoracic, lumbar, or sacral spine.MULTI-AXIAL SCREW FOR 5.5/6.0MM RODS, REF 55840005045

Reason for Recall

This lot has incorrectly been laser marked both on the screw head itself as well as the data carrier tag attached to the screw. The description on face two of the screw head and on the data carrier tag both incorrectly indicate a 5.5 mm diameter screw when the actual screw diameter aligns with 5.0 diameter screw size requirements.

Distribution Pattern

US

Code Information

lot 0648247W